348 VT1422A Command Reference  
[SENSe]
Note
Because of the number of possible strain gage configurations, the driver must 
generate any Strain EU conversion tables and download them to the instrument when 
INITiate is executed. This can cause the time to complete the first INIT command to 
exceed 1 minute on some platforms, notably the Agilent/HP E1405B/06A. 
Subsequent INITs (with no other configuration changes) do not need to regenerate 
EU tables and execute much faster.
Parameters
Comments
Strain measurements require the use of Bridge Completion Signal Conditioning 
Plug-Ons or a Remote Strain Bridge Conditioning Unit.
Bridge Completion SCPs and RSCs provide the strain measurement bridges and 
their excitation voltage sources. <ch_list> specifies the voltage sensing 
channels that are to measure the bridge outputs. Measuring channels on a Bridge 
Completion SCP only returns that SCP's excitation source voltage. 
The VT1422A has five ranges: 0.0625 V dc, 0.25 V dc, 1 V dc, 4 V dc, and 
16 V dc. To select a range, simply specify the range value (for example, 4 selects 
the 4 V dc range). If a value is specified that is larger than one of the first four 
ranges, the VT1422A selects the next higher range (for example, 4.1 selects the 
16 V dc range). Specifying a value larger than 16 generates an error. Specifying 
0 selects the lowest range (0.0625 V dc). Specifying AUTO selects auto range. 
The default range (no range parameter specified) is auto range. 
Command
Bridge Type
Strain SCP and VT1529A/B 
Usage
:FBENding
Full Bending Bridge
Both VT1529A/B and SCPs
:FBPoisson
Full Bending Poisson Bridge
SCPs only
:FPOisson
Full Poisson Bridge
SCPs only
:HBENding
Half Bending Bridge
Both VT1529A/B and SCPs
:HPOisson
Half Poisson Bridge
SCPs only
[:QUARter]
Quarter Bridge (default)
Both VT1529A/B and SCPs.
For VT1529A/B, selects Q350
:Q120
Quarter Bridge 120 
Ω
VT1529A/B only
:Q350
Quarter Bridge 350 
Ω
VT1529A/B only
:USER
Quarter Bridge with user 
installed resistor
VT1529A/B only
Parameter
Name
Parameter
Type
Range of
Values
Default 
Units
range
numeric (float32)
see comments
V dc
ch_list
channel list (string)
100 - 15731
none