346 VT1422A Command Reference  
[SENSe]
[SENSe:]FUNCtion:RESistance
[SENSe:]FUNCtion:RESistance  <excite_current>,[<range>,](@<ch_list>) 
 
links the EU conversion type for resistance and range with the channels specified 
by <ch_list>.
Parameters
Comments
The VT1422A has five ranges: 0.0625 V dc, 0.25 V dc, 1 V dc, 4 V dc, and 
16 V dc. To select a range, simply specify the range value (for example, 4 selects 
the 4V dc range). If a value is specified that is larger than one of the first four 
ranges, the VT1422A selects the next higher range (for example, 4.1 selects the 
16 V dc range). Specifying a value larger than 16 causes an error. Specifying 0 
selects the lowest range (0.0625 V dc). Specifying AUTO selects auto range. 
The default range (no range parameter specified) is auto range.
If using amplifier SCPs, set them first and keep their settings in mind when 
specifying a range setting. For instance, if the expected signal voltage is to be 
approximately 0.1 V dc and the amplifier SCP for that channel has a gain of 8, 
<range> must be set no lower than 1 V dc or an input out-of-range condition 
will exist.
Resistance measurements require the use of Current Source Signal Conditioning 
Plug-Ons.
The <excite_current> parameter (excitation current) does not control the current 
applied to the channel to be measured. The <excite_current> parameter only 
passes the setting of the SCP supplying current to channel to be measured. The 
current must have already been set using the OUTPUT:CURRENT:AMPL 
command. The choices for <excite_current> are 30E-6 (or MIN) and 488E-6 
(or MAX). <excite_current> may be specified in milliamps and microamps.
The *CAL? command calibrates resistance channels based on Current Source 
SCP and Sense Amplifier SCP setup at the time of execution. If SCP settings 
are changed, those channels are no longer calibrated. *CAL? must be executed 
again.
When Accepted: Not while INITiated
Related Commands: OUTP:CURR, *CAL?
*RST Condition: SENSE:FUNC:VOLT AUTO on all channels and bridges set 
to FBEN on all strain channels
Send with VXIplug&play Function: hpe1422_cmd(...)
Usage
FUNC:RES  30ua,(@100,105,107)
Set channels 0, 5, and 7 to convert voltage 
to resistance assuming current source set 
to 30 µA use auto-range (default)
Parameter
Name
Parameter
Type
Range of
Values
Default 
Units
excite_current
 discrete(string)
30E-6 | 488E-6 | MIN | MAX
Amps
range
 numeric (float32)
see first comment
V dc
ch_list
channel list (string)
100 - 163
none