SITRANS LC 500 – INSTRUCTION MANUAL  
7ML19985GE01
mmmm
m
Oper
at
ion & Appl
icat
ion
In a non-conductive or irregular tank
Where the vessel contents are non-conductive: 
• a reference electrode parallel to the measurement electrode is required 
• the reference electrode must be grounded to the instrument
• a stilling well can form the reference electrode.
Where the vessel contents are conductive:
• the interface between the contents and the electrode insulation acts as the 
reference electrode
• a connection from the vessel contents to the instrument ground is required
• a stilling well can provide a means of connecting the contents to the instrument 
ground. 
The stilling well
The stilling well is a metal tube concentric with the electrode, with vent openings to 
facilitate level equalization. Its diameter is somewhat larger than that of the electrode, 
depending on the application. The stilling well can either be integral to the SITRANS LC 500, 
or it may be part of the tank
1
The SITRANS LC 500 electrode
The SITRANS LC 500 electrode, comprising a measurement section and an active shield 
section, is the primary sensor of the system. It supplies the electrical capacitance value 
of the measurement section relative to the environment (tank wall or stilling well). 
The SITRANS LC 500 patented Active-Shield Technology electrically isolates the 
measurement section and prevents any non-measurement capacitance from interfering 
with the measurement. (Capacitance changes could result from uncontrolled variations 
occurring in the connection cable, process connection, and non-active parts of the 
probe). This results in a better ratio of initial capacitance to total capacitance, resulting in 
higher accuracy.
1.
The tank wall, or the stilling well if it is part of the tank, must be grounded.