SITRANS LC 500 – INSTRUCTION MANUAL 
Page 11
mm
mmm
Oper
at
ion & Applicat
ion
The measurement is further protected from interference by a buffer, which applies the 
frequency signal from the measurement section to the active shield section. This effectively 
eliminates any electrical potential difference between the shield and the measurement 
section and prevents additional changes in capacitance occurring.
The relative lengths of the measurement section and active shield section can be 
specified to suit a particular application. If the measured range will be short relative to 
the total length of the electrode, specify a short measurement section. This increases the 
achievable resolution of the measurement, since any change in level will be greater 
relative to the length of the measurement section.
Measuring-
Circuit
Measuring-
Circuit
Conventional Capacitance 
Measurement
SITRANS LC 500 
with Active Shield
C1
C2
C3
R  =  (C1 + C2 + C3) + Ca
(C1 + C2 + C3) + Ca + Cm
R   =  Ca
Ca + Cm
Ca
Ca
Cm
Cm
R  = Ratio between initial 
capacitance and total 
capacitance
Ca = Initial capacitance (air)
Cm = Capacitance Increase 
(product)
C1 = 
Capacitance 
connection 
point
C2  = Capacitance connection 
cable
C3 = Capacitance 
Process 
connection (includes active 
part)
 frequency  (f) 
K
capacitance (C)
active shield
active 
measurement 
section
probe seal 
(inactive)
100%
0%
SITRANS 
LC 500
buffer
empty 
tank
full 
tank