Siemens syngo CT 2005C 사용자 설명서

다운로드
페이지 321
Thorax
99
 
Hints in General
• Topogram: TOP, 512 mm.
• Patient positioning: Patient lying in supine position, 
arms positioned comfortably above the head in the 
head-arm rest, lower legs supported.
• Contrast medium administration: in general, IV 
injections are employed in all mediastinal examina-
tions, but not in routine high resolution studies of 
diffuse interstitial lung diseases. An IV contrast 
medium injection improves the vascular opacifica-
tion and facilitates the visualization of the lesions, 
lymph nodes and the vessels.
• Stasis of contrast medium in the arm & superior vena 
cava often result in high density streak artifacts 
either in the region of the aortic arch or in the region 
of the subclavian vein. A caudo-cranial (bottom to 
top) scanning direction should be used to reduce this 
artifact – by simply acquiring the data in this region 
at the later phase of the spiral scan. In addition, if the 
patient cannot hold his/her breath for the duration of 
the entire scan, breathing motion will be less appar-
ent in the apex than in the lower lobes.
• CARE Bolus may be used to optimize the bolus tim-
ing. Set the ROI for monitoring scan in the aorta at 
the level of the diaphragm with triggering threshold 
of 120 HU, or use manual triggering.
C2-025.630.01.01.02_APPLICATIONGUIDE_SPIRIT.book  Page 99  Friday, April 8, 2005  9:55 AM