Teledyne Camera Accessories T100 사용자 설명서
Communications Setup and Operation
Teledyne API - T100 UV Fluorescence SO2 Analyzer
152
To assign or reset the status flag bit assignments, access the COMMUNICATIONS
Menu (refer to Figure 5-13), then press:
Menu (refer to Figure 5-13), then press:
SETUP X.
HESSEN STATUS FLAGS
<SET
SET>
EDIT
EXIT
SETUP X.X
COMMUNICATIONS MENU
ID INET
HESN
COM1
COM2
EXIT
SETUP X.
PMT DET WARNING: 0002
PREV NEXT
EDIT
PRNT
EXIT
Repeat pressing SET> until …
SETUP X.
SYSTEM RESET: 0000
PREV NEXT
EDIT
PRNT
EXIT
Press <CH and
CH>
to move the
[ ]
cursor left
and right along
the bit string.
Repeat pressing NEXT or PREV until the desired
message flag is displayed. Refer to Table 7-15.
For Example …
ENTR accepts the new
settings
EXIT ignores the new
settings
SETUP X.
SYSTEM RESET: [0]000
<CH CH>
[0] ENTR EXIT
Press [?]repeatedly to cycle through the available character set: 0-9
Note: Values of A-F can also be set but are meaningless.
Figure 6-10:
COMM – Status Flag Bit Assignment
6.6.2.7. INSTRUMENT ID
Each instrument on a Hessen Protocol network must have a unique identifier (ID
number). Refer to Section 5.7.1 for information and to customize the ID of each.
number). Refer to Section 5.7.1 for information and to customize the ID of each.