Communications Setup and Operation 
Teledyne API - T100 UV Fluorescence SO2 Analyzer  
152 
To assign or reset the status flag bit assignments, access the COMMUNICATIONS 
Menu (refer to Figure 5-13), then press: 
 
 
 
 
SETUP X.       
HESSEN STATUS FLAGS                       
 
 <SET
  
 
SET>
    EDIT
 
                                  
 EXIT
 
 
 
 
SETUP X.X 
COMMUNICATIONS MENU
 
  ID    INET    
HESN
    
COM1
    
 COM2
   
                     EXIT
 
SETUP X.      
   PMT DET WARNING: 0002                      
 
PREV   NEXT
 
                          
EDIT    
 PRNT
   EXIT
 
Repeat pressing SET> until … 
SETUP X.      
    SYSTEM RESET: 0000                        
 
PREV   NEXT
 
                          
EDIT    
 PRNT
   EXIT
 
Press  <CH and 
CH> 
to move the 
[  ] 
cursor left 
and right along 
the bit string. 
Repeat pressing NEXT or  PREV  until the desired 
message flag is displayed. Refer to Table 7-15.  
For Example … 
ENTR accepts the new 
settings  
EXIT ignores the new 
settings  
SETUP X.      
    SYSTEM RESET: [0]000                        
 
 <CH    CH>
 
                          
[0]        ENTR   EXIT
 
Press [?]repeatedly to cycle through the available character set:  0-9
Note: Values of A-F can also be set but are meaningless. 
 
Figure 6-10: 
COMM – Status Flag Bit Assignment 
6.6.2.7. INSTRUMENT ID 
Each instrument on a Hessen Protocol network must have a unique identifier (ID 
number).  Refer to Section 5.7.1 for information and to customize the ID of each. 