Teledyne Camera Accessories T100 사용자 설명서

다운로드
페이지 422
Data Acquisition System (DAS) and APICOM 
Teledyne API - T100 UV Fluorescence SO2 Analyzer  
154 
Table 7-1: Front Panel LED Status Indicators for DAS 
LED STATE 
DAS STATUS 
Off 
System is in calibration mode. Data logging can be enabled or disabled for this mode. 
Calibration data are typically stored at the end of calibration periods, concentration data are 
typically not sampled, diagnostic data should be collected. 
Blinking 
Instrument is in hold-off mode, a short period after the system exits calibrations. DAS 
channels can be enabled or disabled for this period. Concentration data are typically disabled 
whereas diagnostic should be collected. 
On Sampling 
normally. 
 
Note 
The DAS can be disabled only by disabling or deleting its individual data 
channels. 
 
7.1. DAS STRUCTURE 
The DAS is designed around the feature of a “record”. A record is a single data point of 
one parameter, stored in one (or more) data channels and generated by one of several 
triggering event. The entire DAS configuration is stored in a script, which can be edited 
from the front panel or downloaded, edited and uploaded to the instrument in form of a 
string of plain-text lines through the communication ports. 
DAS data are defined by the PARAMETER type and are stored through different 
triggering EVENTS in data CHANNELS, which relate triggering events to data 
parameters and define certain operational functions related to the recording and 
reporting of the data. 
7.1.1. DAS CHANNELS 
The key to the flexibility of the DAS is its ability to store a large number of 
combinations of triggering events and data parameters in the form of data channels. 
Users may create up to 20 data channels and each channel can contain one or more 
parameters. For each channel one triggering event is selected and up to 50 data 
parameters, which can be the same or different between channels. Each data channel has 
several properties that define the structure of the channel and allow the user to make 
operational decisions regarding the channel (refer to Table 7-2). 
06807C DCN6650