SECTION 2—ASSEMBLY
12
Assembling the Electric Actuator to the Boom
NOTE: For this procedure, refer to FIGURE 2.2.
1. Remove the shoulder bolt, washer and 
nut from the mounting bracket on the 
boom assembly.
2. Unpack the pinch guard from the patient 
lift carton.
NOTE: The bottom of the electric actuator 
assembly will already be assembled to the mast 
mounting bracket.
3. Cut the plastic‐wrap that secures the 
boom and mast together.
4. Lift‐up on the boom and place it on your 
left shoulder. 
5. Let the actuator rest on your right‐side of your chest and rotate the shaft extension of the 
actuator assembly until it lines‐up with the mounting holes in the boom assembly.
6. Place the pinch guard over the shaft extension of the actuator.
7. Align the holes of the boom assembly mounting bracket with those of the actuator and 
insert the bolt. Secure with nut.
NOTE: Be sure that the bolt is completely through the holes of the boom assembly mounting bracket and 
the actuator assembly. The boom assembly will pivot easily if the mounting hardware is aligned properly 
when the boom assembly is secured to the mast.
CAUTION
DO NOT overtighten the nut and bolt. This damages the mounting bracket.
8. Plug in the electrical connectors from the electric actuator to the bottom of the battery 
assembly.
9. Plug the pendant control into the bottom of the battery assembly.
NOTE: The bottom of the actuator assembly will already be assembled to the mast mounting bracket.
Pinch Guard 
NOTE: Turn 90°
Mounting Bracket