Invacare Electric Portable Patient Lift RPA600-1E 사용자 설명서
SECTION 2—ASSEMBLY
12
Assembling the Electric Actuator to the Boom
NOTE: For this procedure, refer to FIGURE 2.2.
1. Remove the shoulder bolt, washer and
nut from the mounting bracket on the
boom assembly.
boom assembly.
2. Unpack the pinch guard from the patient
lift carton.
NOTE: The bottom of the electric actuator
assembly will already be assembled to the mast
mounting bracket.
assembly will already be assembled to the mast
mounting bracket.
3. Cut the plastic‐wrap that secures the
boom and mast together.
4. Lift‐up on the boom and place it on your
left shoulder.
5. Let the actuator rest on your right‐side of your chest and rotate the shaft extension of the
actuator assembly until it lines‐up with the mounting holes in the boom assembly.
6. Place the pinch guard over the shaft extension of the actuator.
7. Align the holes of the boom assembly mounting bracket with those of the actuator and
insert the bolt. Secure with nut.
NOTE: Be sure that the bolt is completely through the holes of the boom assembly mounting bracket and
the actuator assembly. The boom assembly will pivot easily if the mounting hardware is aligned properly
when the boom assembly is secured to the mast.
the actuator assembly. The boom assembly will pivot easily if the mounting hardware is aligned properly
when the boom assembly is secured to the mast.
CAUTION
DO NOT overtighten the nut and bolt. This damages the mounting bracket.
8. Plug in the electrical connectors from the electric actuator to the bottom of the battery
assembly.
9. Plug the pendant control into the bottom of the battery assembly.
NOTE: The bottom of the actuator assembly will already be assembled to the mast mounting bracket.
Pinch Guard
NOTE: Turn 90°
Mounting Bracket