Measuring Insertion Losses
To make accurate power and receiver measurements, the signal loss 
through the cables or other devices used in your test setup must be 
known and entered into the Test Set’s INSTRUMENT CONFIGURE 
screen to compensate for these losses. 
Signal losses are measured using a  built-in automated routine that 
runs on the Test Set’s IBASIC controller. Losses can be calibrated at a 
discrete (single) frequency or over a frequency range. This is one 
routine included in a set of utility procedures called the RF TOOLS.
During the test, a calibrated signal goes through two 6-dB 
attentuators/pads (such as Mini-Circuits model NAT-6-60) and a short 
type-N male-to-male cable to establish a known reference point. The 
Test Set then prompts you to connect the device under test to measure 
the additional loss through that device. (The pads and cable are not 
part of the standard equipment shipped with the Test Set.)
 shows how to load and run the RFTOOLS routines and 
select the desired insertion loss test.
Figure 3-1
Loading and Running the Insertion Loss Test
1
Select ROM.
2
Select RFTOOLS.
3
Select Run Test.
When the screen below appears.......
...use the knob to move the cursor and 
select either Discrete Freq Insertion 
Loss
 or Swept Insertion Loss. When 
run, the test prompts you to enter the 
necessary test frequency information 
and displays setup diagrams. Make the 
indicated connections and follow 
instructions as you are prompted by the 
program. 