Gossen R2601 사용자 설명서

다운로드
페이지 40
GOSSEN METRAWATT GMBH
22
3.6 Send data to R2600/01
This communication makes it possible to set all parameters, configurations and operating states which can be changed via operation. Thereby, the data 
is individually addressed per parameter index. See section 4 for the complete list of all parameter indices. There is no protection against overwriting. 
The settings of the DIP switches for disabling of the configuration and/or parameter setting are of no importance. Also the start of optimizing cannot be 
disabled.
The R2600/01 checks the transmitted value for its setting range. Should it be beyond its permissible range, it is not stored, the bit "impermissible 
value" is set in error status word 1, and the "service request" bit is set in the acknowledgement short set.
Same as with manual operation of the equipment, it must be noted, that a complete configuration must be made before parameters are set: That is, the 
"control instructions" and "equipment specifications" must first be transmitted which influence the presentation of the "temperature parameters" (see 
4.3 and 4.4).
3.6.1
Sending an equipment specification
The parameter index lies between 30h and 3Fh. Thus the characters "from/to channel" and "receipt No." are omitted in the long set.
Example: Setting the thermocouple type "K" on the equipment having the address = 0 (see 4.4).
Request from master (long set, see 2.3):
68h
05h
05h
68h
00h
69
33h
02h
00h
9Eh
16h
FF
PI
PS
Data block