MR810 Technical Manual Revision C  REF 185042601
 
21 
 
6
6.4 R
EPLACING THE 
PCB 
Table 5 : Spare PCB part numbers 
Model 
Supply Voltage 
PCB 
Part Number 
MR810A-- 
230 V~ 
043 042 232 
MR810J-- 
115 V~ 
043 042 233 
MR810G-- 
100 V~ 
043 042 234 
 
1.  Open the case and remove the PCB (refer to § 6.2). 
2.  Cut the cable tie holding the phase and neutral conductors to the PCB (refer to Figure 6.4). 
3.  Slide the power cable and heater-wire adaptor clamp off the transformer support and lift away. 
4.  Unscrew the power cable (phase, neutral, and earth), and the heater plate earth connection. 
5.  Detach the heater-wire adaptor connector, the heater plate connector and the two heater plate thermistor 
connectors (refer to Figure 6.4). 
6.  Unpack the replacement PCB and check that it is the correct voltage model from Table 5 above. 
7.  Reconnect the power cable, secure the phase and neutral wires to the PCB using a small (2.5 mm wide) 
cable tie. 
8.  Screw the heater plate protective earth into the protective earth terminal block. 
9.  Reconnect the heater-wire adaptor, heater plate harness and two thermistors to their respective 
locations.  
Note: the primary and secondary thermistors can be connected to either location on the PCB.  
10.  Slide the power cable and heater-wire adaptor clamp onto the transformer of the new PCB. 
11.  Reinstall the PCB, and close the case (refer to § 6.8). 
 
 
 
Figure 6.4: Harness and power cable connections 
 
Heater Plate 
protective earth 
Mains protective 
earth 
Heater plate 
connection 
 Cable tie 
Primary 
thermistor 
connection 
Secondary thermistor 
connection 
Heater wire adaptor 
connection