MR810 Technical Manual Rev C  REF 185042601
 
 
 
6 
6.5 S
ERVICING THE 
H
EATER 
P
LATE
 
Table 6: MR810 Screw Torque Settings 
Exploded 
Diagram 
Reference 
Screw 
Application 
Driver 
Torque 
(in/oz) 
Torque 
(Nm) 
Figure 8.2: 9,11 
Screw M4 x 8 Pan Phil 
Taptite 
Heater Plate Earth, 
Heater Plate Element 
Philips #2 
205–234  1.45-1.65 
Figure 8.2: 3,6 
Screw M3 x 5 Pan Phil 
Taptite 
Thermostat, 
Thermistor 
Philips #1 
163-177 
1.15-1.25 
Figure 8.1: 5 
Screw M3 x 5 Pan Phil 
Taptite 
Heater Plate Stand-off  
Philips #2 
85– 113 
0.6 – 0.8 
Figure 8.1: 1,2 
Screw 6 x 53 Pan Phil 
TY25 ZP 
MR810 PCB, 
Case screw 
Philips #2 
99 –127 
0.7 – 0.9 
6.5.1 R
ESETTING THE 
T
HERMAL 
C
UT
-O
UT
 
NOTE: If the heater plate is still hot, allow it to cool sufficiently before attempting to reset the thermal cut-
out. 
1.  Disconnect the power supply and open the case (refer to § 6.2). 
2.  With the humidifier upside down, the thermal cut-out can be reset by depressing the red thermal cut-out 
reset button through the hole in the PCB (Figure 6.5), using a pin of at least 60 mm long and diameter 
of approximately 4-5 mm.  If the thermal cut-out “clicks” when pressed then the cut-out had been 
tripped.  
3.  Close the case (refer to § 6.8.2). 
 
Figure 6.5: Location of the thermal cut out reset button 
 
Thermal cut-out 
reset