6. Press Enter to confirm or Esc to cancel.  If confirmed, the controller 
calibrates the sensor’s Zero.  This takes only a few moments and 
the display returns to the default calibration screen. 
 
Note: Never calibrate any unit’s Zero with ambient air. Always use 
Nitrogen (N
2
) at the calibration port to calibrate the Zero. 
 
7. To calibrate the device, scroll to Set Span** and change the span 
gas calibration value using this procedure;  
 
 
 
a.   Using the right arrow, move the cursor to xxx  PPM (span 
value field).  Press Enter to select the field (it is editable 
when flashing). 
b.   Use the up or down arrows to increase or decrease the 
value, press Enter to validate the new value. 
c.   Move the cursor back to Set Span and press Enter to start 
the calibration.  
 
 
 
 
The device Span is being calibrated.  The screen will display the 
device’s status as “In calib...” until the calibration is complete. 
 
**When selecting Set Span, make sure that the device has been 
supplied with the appropriate calibration gas before and during the 
calibration process.   
61