The Tests menu provides four main options, divided between two 
screens. Each of these options offers different capabilities. 
 
The first screen presents three options: 
 
 
 
Test sequence:  Enables each output to be activated and validates 
operation of each controller keypad buttons, display 
pixels, and various communication protocols. 
Start Sim  
Sequence: 
This options starts or stops the Simulation mode, 
which allows a simulation of a gas concentration 
over an associated scale range on all transmitters.  
The simulated gas concentration values are local (on 
the controller) and do not affect logging functions. 
(Events will be activated for the simulation but 
detection devices are not affected.) 
Maximum load: 
Activates all controller components 
 
The second screen option is “Oprt Mode”, which offers three separate 
operation settings: Normal, Single Tx or Debug. 
 
 
 
Normal 
Normal controller operation mode 
Single Tx: 
Activates the polling mode on a single transmitter. 
Debug: 
Activates the service mode to perform a calibration 
and to test Events without triggering actions. 
63