Model NGA2000 HFID
 
1-5 SPECIFICATIONS 
a. General 
Measurement Species ....................Total hydrocarbons 
Ranges (H
2
/He Fuel) 
Low range ............................0 to 10 ppm, CH
4
, through 0 to 1%, CH
at an oven setpoint between 
113
°C and 191°C 
High range ...........................0 to 50 ppm, CH
4
, through 0 to <5%, CH
4
 at an oven setpoint 
between 113
°C and 191°C 
Analysis Temperature .....................Adjustable from 200
o
F to 400
o
F (93
o
C to 204
o
C), maintained within 
±11
o
F (±6
o
C) from the setpoint. 
Repeatability ....................................
≤1% of fullscale for successive identical samples, at a constant 
temperature, sample flow and fuel, burner air, regulated air and 
sample pressures 
Min. Detectable Level......................0.10 ppm, CH
4
 
Noise ................................................
≤1% of fullscale, peak to peak 
Linearity............................................
≤±1% of fullscale, ≤±2% of data point (must be above the minimum 
detectable level) 
Response Time ...............................
≤1.5 sec., 0% to 90% of fullscale 
Drift 
Zero ..................................... 
≤ ±1% of fullscale/24 hours at constant temperature, sample flow, 
hydrocarbon concentration of supply gases, and fuel, burner air, 
regulated air and sample pressures. 
Span .................................... 
≤ ±1% of fullscale/24 hours at constant temperature, sample flow, 
hydrocarbon concentration of supply gases, and fuel, burner air, 
regulated air and sample pressures. 
Effect of Temperature....................
≤±2% of fullscale for any ambient temperature change of 10
o
C and 
rate of change less than 10
o
C/hr. 
Operating Temperature ................. 59
°F to 95°F (15°C to 35°C) 
Power Requirements.......................+24 VDC ±5%, 120 W max. direct to Analyzer Module 
Ripple and Noise .................<100 mV pp 