Emerson Process Management NGA2000 HFID Manual Do Utilizador

Página de 76
Instruction Manual 
748297-E 
June 2003 
Rosemount Analytical Inc.    A Division of Emerson Process Management                                   Description and Specifications 
1-5
Model NGA2000 HFID 
b. Gas 
Requirements 
Sample .......................................... Non-flammable, below 100% of LEL 
Flow rate ..............................1.0 to 2.5 L/min. 
Supply pressure...................345 to 620 hPa-gauge (5 to 9 psig) 
Temperature ........................110
o
C to 230
o
C (230
o
F to 446
o
F), <20
o
C variance/24 hours, <10
o
variance/hr. 
Particulates ..........................filtered to <2 microns 
Dewpoint..............................<15
o
C below the setpoint 
Regulated Air................................. Instrument air or nitrogen 
Flow rate ..............................2 to 4 L/min. 
THC .....................................
≤2 ppm, CH
4
 
Supply pressure...................689 to 1723 hPa-gauge (10 to 25 psig) 
Particulates ..........................filtered to <2 microns 
Purge Air:....................................... Instrument air, nitrogen or other nonflammable gas (refer to 
ANSI/NFPA 496 for the requirements for the Protective Gas Sys-
tem) 
Flow rate:  ............................16 to 18 L/min. 
Supply pressure:..................689 to 1378 hPa-gauge (10 to 20 psig) 
Fuel Gas ........................................ Premixed 40% hydrogen and 60% helium 
Flow rate ..............................80 to 100 ml/min 
THC .....................................
≤0.5 ppm, CH
4
 
Supply pressure...................3101 to 3446 hPa-gauge (45 to 50 psig) 
 
WARNING 
EXPLOSION HAZARD 
Do not use pure hydrogen fuel.  An explosion resulting in severe personal injury or death could oc-
cur. 
Burner Air ...................................... Zero-grade air 
Flow rate ..............................350 to 400 mL/min. 
THC .....................................
≤1 ppm, CH
4
 
Supply pressure...................1723 to 3446 hPa-gauge (25 to 50 psig) 
 
c. Physical 
Case Classification ........................ General purpose for installation in weather-protected area 
Maximum Separation .................... 1600m (1 mile) from Analyzer Module to Platform 
Materials in Contact With Sample Stainless steel and glass-filled teflon 
Dimensions.................................... See Outline and Mounting Dimensions, Figure 2-5 on page 2-7 
Weight ........................................... 15.9 kg (35 lbs.) 
Mounting........................................ Horizontally, custom-installed in a panel