10.13
Checking process variables
Micro Motion suggests that you make a record of the process variables listed below, under normal
operating conditions. This will help you recognize when the process variables are unusually high or
low.
operating conditions. This will help you recognize when the process variables are unusually high or
low.
•
Flow rate
•
Density
•
Temperature
•
Tube frequency
•
Pickoff voltage
•
Drive gain
For troubleshooting, check the process variables under both normal flow and tubes-full no-flow
conditions. Except for flow rate, you should see little or no change between flow and no-flow
conditions. If you see a significant difference, record the values and contact Micro Motion customer
service for assistance.
conditions. Except for flow rate, you should see little or no change between flow and no-flow
conditions. If you see a significant difference, record the values and contact Micro Motion customer
service for assistance.
Unusual values for process variables may indicate a variety of different problems. Table 10-4 lists
several possible problems and suggested remedies.
several possible problems and suggested remedies.
A106
Burst Mode
Enabled
Enabled
Device is in HART burst
mode.
mode.
• No action required.
• If desired, reconfigure the alarm severity to Ignore (see
• If desired, reconfigure the alarm severity to Ignore (see
Burst Mode
Enabled/AI or AO
Simulate Active
Enabled/AI or AO
Simulate Active
A107
Power Reset
Occurred
Occurred
The transmitter has been
restarted.
restarted.
• No action required.
• If desired, reconfigure the alarm severity to Ignore (see
• If desired, reconfigure the alarm severity to Ignore (see
Power Reset
Occurred
Occurred
A132
Simulation Mode
Active
Active
Simulation mode is
enabled.
enabled.
• Disable sensor simulation. See Section 7.5.
Simulation Mode
Active
Active
Table 10-4
Process variables problems and remedies
Symptom
Cause
Suggested remedy
Steady non-zero flow rate under
no-flow conditions
no-flow conditions
Misaligned piping (especially in new
installations)
installations)
• Correct the piping.
Open or leaking valve
• Check or correct the valve
mechanism.
Bad sensor zero
• Rezero the flowmeter. See
Table 10-3
Status alarms and remedies continued
Alarm
code
code
Communicator
Cause
Suggested remedy
ProLink II