SECTION 5 —OPERATING INSTRUCTIONS
19
Recommended Guidelines for Optimum 
Performance
NOTE: Ensure that your concentrator is at least three inches away 
from walls, draperies or furniture to assure sufficient air flow. Avoid 
deep pile carpets and heaters, radiators or hot air registers.
Plug in Power Cord
1. Plug in power cord to an electrical outlet.
Temperature:
50°F - 95°F (10°C - 35°C)
Electrical:
No extension cords.
Placement:
No closer than three inches from the wall, 
furniture, draperies, or similar surfaces.
Tubing and Cannula:
7 ft cannula with a maximum 50 ft of 
crush-proof Tubing (DO NOT pinch).
Platinum 10: Recommended use up to 50 ft 
high flow tubing with high flow cannula at all 
flow rates.
Environment:
Smoke and soot-free. No confined spaces 
(Example: No closets).
Relative Humidity:
20 to 60%
Time of Operation:
Up to 24 hours per day.
Flow Rate 
Platinum 5 and XL: From 0.5 L/min. to 5 
L/min. For flow rates less than 1 L/min., use 
the Pediatric Flowmeter Accessory.
Platinum 10: From 2 L/min. to 10 L/min
Minimum Operating Time:
30 Minutes