SECTION 5 —OPERATING INSTRUCTIONS
20
Connect Humidifier (If So Prescribed)
NOTE: For this procedure, refer to FIGURE 5.1 - FIGURE 5.4.
NOTE: Fill humidifier with distilled water to the level indicated by 
the manufacturer. (For Platinum 5, XL - recommended humidifiers 
are: Invacare 3260, 003-40, or 006-40. For Platinum 10, 
recommended humidifiers are high flow: 7900-10, 7900-25).
 WARNING
DO NOT overfill humidifier.
Replace the humidifier cap and securely tighten. 
DO NOT reverse the oxygen input and output 
connections. Water from the humidifier bottle will travel 
through the cannula back to the patient.
FIGURE 5.1    Filling the Humidifier
2. Place the humidifier bottle in the humidifier 
compartment.
FIGURE 5.2    Humidifier Compartment
Humidifier Bottle
Humidifier Bottle