SECTION 5 —OPERATING INSTRUCTIONS
25
SensO
2
 Oxygen Purity Indicator (If Your 
Unit Features the O
2
 Sensor Feature)
This feature monitors the purity of oxygen generated by the 
oxygen concentrator. If purity falls below factory preset 
standards, indicator lights on the control panel will 
illuminate.
Initial Startup of the Concentrator
NOTE: Concentrator may be used during the initial start warm-up 
time (approximately 30 min.) while waiting for the O
2
 purity to 
reach maximum.
When the unit is turned on, the GREEN light will come on 
(System OK/O
2
 greater than 85%). After five minutes, the 
oxygen sensor will be operating normally and will control the 
indicator lights depending on oxygen concentration values. 
Explanation of Oxygen Purity Indicator 
Lights
NOTE: For this procedure, refer to FIGURE 5.7 on page 26.
GREEN light (O
2
) - Normal Operation.
YELLOW light ( ) - Call supplier IMMEDIATELY. You may 
continue to use the concentrator unless instructed otherwise 
by your supplier. Be certain that backup oxygen is nearby.
RED light ( )- Total unit shutdown. Switch IMMEDIATELY 
to backup oxygen supply. Call supplier IMMEDIATELY.
GREEN light - with YELLOW light flashing - Call supplier 
IMMEDIATELY. Oxygen sensor malfunctioning; you may 
continue to use the concentrator.