SECTION 6 —MAINTENANCE
27
SECTION 6 — MAINTENANCE
NOTE: The Invacare concentrators are specifically designed to 
minimize routine preventive maintenance. In places with high dust 
or soot levels, maintenance may need to be performed more often. 
Only authorized or factory-trained personnel should perform 
preventive maintenance on the concentrator. Power should be 
disconnected before beginning preventive maintenance on the 
Invacare concentrator.
NOTE: At a minimum, preventive maintenance MUST be 
performed according to the maintenance record guidelines. Refer to 
Routine Maintenance
NOTE: The following routine maintenance should be performed 
more frequently. See recommended intervals in each section and 
preventive maintenance record. 
Cleaning the Cabinet Filter
NOTE: For this procedure, refer to FIGURE 6.1. 
NOTE: There are two cabinet filters one located on each side of the 
cabinet.
1. Remove each filter and clean at least once a week 
depending on environmental conditions.
NOTE: Environmental conditions that may require more frequent 
cleaning of the filters include but are not limited to: high dust, 
smoking, air pollutants, etc.
2. Clean the cabinet filters with a vacuum cleaner or wash in 
warm soapy water and rinse thoroughly.
3. Dry the filters thoroughly before reinstallation.