Pacing
8-5
8
4. Adjust the rate to the desired number of paced pulses per minute (ppm). 
Press up, 
, or down, 
, on 
 to increase or decrease the num-
ber of paced pulses per minute.
5. Press 
 to start pacing.
The message 
3DFLQJ indicates that paced pulses are being delivered in the 
selected mode at the rate and output level displayed.
Figure 8-2:  Pacing Display
NOTE
Pacing will not start if there is a problem with the multifunction defib elec-
trode pads connections or, in demand mode, a problem with the ECG monitor-
ing electrodes connections. Should a problem exist, a system message is 
displayed. Be attentive to system messages.