Changing Pacing Modes
8-6
Pacing
6. Increase the output until cardiac capture occurs.
Press up 
 on 
 to increase the output in increments of 10 mA. 
7. Decrease the output to the lowest level that still maintains capture.
Press 
 on 
 to decrease the output in increments of 5 mA. 
Press 
 to stop pacing. Press 
 to exit the pacing function. The 
green LED next to the button goes out, indicating pacing is no longer active.
Changing Pacing Modes
If paced pulses are being delivered, you must stop pacing before changing the 
pacing mode:
1. Press 
 to stop pacing.
2. Press 
 to change the mode.
3. Adjust the rate, if needed.
4. Press 
 to resume pacing.
5. Adjust the output, as needed to obtain capture.
Defibrillating During Pacing
If the patient must be defibrillated during pacing, follow the procedure for 
defibrillating in Manual Mode on page 6-4. 
Pacing is automatically turned off when you charge the defibrillator and the 
pacing dialogue box is removed from the display. After a shock, pacing 
remains off.
To resume pacing, refer to “Pacing” on page 8-4. When you resume, the rate, 
mode, and output settings start at the settings selected prior to defibrillation.
Output
Output
Start
Stop
Pacer
Start
Stop
Mode
Start