Siemens LR 460 Manual Do Utilizador
Page 40
SITRANS LR 460 – INSTRUCTION MANUAL
7ML19985JM01
mmmmm
S
IMA
T
IC PD
M
Changing parameter settings via SIMATIC PDM
1.
Launch SIMATIC PDM, connect to SITRANS LR 460, and upload data from the
device.
device.
2.
Adjust parameter values in the parameter value field then press Enter. The status
fields read Changed.
3.
Open the Device menu, click on Download to device, then use File – Save, to save
parameter settings. The status fields are cleared.
Configuring a new device
Device Description (DD)
You will need the DD for SIMATIC PDM version 6.0 with SP2. You can locate the DD in
Device Catalog, under Sensors/Level/Echo/Siemens Milltronics/SITRANS LR 460. If you do
not see SITRANS LR 460 under Siemens Milltronics, you can download the DD from the
product page of our website at: https://pia.khe.siemens.com/index.asp?Nr=14655, under
Downloads.
Device Catalog, under Sensors/Level/Echo/Siemens Milltronics/SITRANS LR 460. If you do
not see SITRANS LR 460 under Siemens Milltronics, you can download the DD from the
product page of our website at: https://pia.khe.siemens.com/index.asp?Nr=14655, under
Downloads.
Save the files to your computer, and extract the zipped file to an easily accessed location.
Launch SIMATIC PDM – Manager Device Catalog, browse to the unzipped DD file and
select it.
Launch SIMATIC PDM – Manager Device Catalog, browse to the unzipped DD file and
select it.
1.
Set Address (default for PROFIBUS PA is 126; for HART is 0).
• Point the handheld programmer at the display then press Mode
• Point the handheld programmer at the display then press Mode
to
activate
PROGRAM mode, menu item 1.0.
• Press Down ARROW
, Right ARROW ,
Right ARROW
to navigate to
Address (menu item 2.1.2).
• Press Right ARROW
to open Edit mode: the PROGRAM icon will flash.
• If required, key in a new value and press Right ARROW
to accept it. The LCD
displays the new value and the PROGRAM icon disappears.
Note:
value
fields