Siemens LR 460 사용자 설명서

다운로드
페이지 123
Page 40
SITRANS LR 460 – INSTRUCTION MANUAL
7ML19985JM01
mmmmm
S
IMA
T
IC PD
M
Changing parameter settings via SIMATIC PDM
1.
Launch SIMATIC PDM, connect to SITRANS LR 460, and upload data from the 
device.
2.
Adjust parameter values in the parameter value field then press Enter. The status 
fields read Changed.
3.
Open the Device menu, click on Download to device, then use File – Save, to save 
parameter settings. The status fields are cleared.
Configuring a new device
Device Description (DD)
You will need the DD for SIMATIC PDM version 6.0 with SP2. You can locate the DD in 
Device Catalog, under Sensors/Level/Echo/Siemens Milltronics/SITRANS LR 460. If you do 
not see SITRANS LR 460 under Siemens Milltronics, you can download the DD from the 
product page of our website at: https://pia.khe.siemens.com/index.asp?Nr=14655, under 
Downloads. 
Save the files to your computer, and extract the zipped file to an easily accessed location. 
Launch SIMATIC PDM – Manager Device Catalog, browse to the unzipped DD file and 
select it. 
1.
Set Address (default for PROFIBUS PA is 126; for HART is 0).
• Point the handheld programmer at the display then press Mode  
to 
activate 
PROGRAM mode, menu item 1.0. 
• Press Down ARROW 
, Right ARROW , 
Right ARROW 
 to navigate to 
Address (menu item 2.1.2). 
• Press Right ARROW 
 to open Edit mode: the PROGRAM icon   will flash.
• If required, key in a new value and press Right ARROW 
 to accept it. The LCD 
displays the new value and the PROGRAM icon disappears.
Note: 
For a complete list of parameters, see 
value 
fields