Operating the T703 Calibrator 
Teledyne API T703 Calibrator Operation Manual 
100  
 
6.12.5. PROGRAMMING NEW LEVELS 
To begin programming a new LEVEL find the LVL submenu by pressing: 
 
CHOOSE ACTION Submenu
SETUP X.X      
ACTION TO PERFORM:GENERATE
PREV NEXT
ENTR
EXIT
STANDBY            ACT =STANDBY
<TST
TST>
GEN STBY SEQ
SETUP
Make sure that the 
calibrator is in standby 
mode.
SETUP X.X        
PRIMARY SETUP MENU
O3
LEVL
SEQ
CFG
CLK PASS
MORE
EXIT
SETUP X.X               
END OF LEVELS
INS
PRNT
EXIT
SETUP X.X  
[LEVEL ID] 
[Gas/Conc.], [Status Block Set]
PREV NEXT
INS
DEL EDIT
PRNT
EXIT
Deletes the LEVEL shown
in the message field
Edits the LEVEL shown in 
the message field
Scrolls back and forth between 
existing LEVELS
This display only appears if there are no LEVELs currently 
programmed into the calibrator.
OTHERWISE ...
Toggle to scroll though the available instructions: 
GENERATE & MANUAL
 
 
07223B DCN6378