Teledyne MODEL T703 Manual Do Utilizador

Página de 287
Operating the T703 Calibrator 
Teledyne API T703 Calibrator Operation Manual 
102  
 
6.12.5.2.  Editing or Deleting a LEVEL   
To edit or delete an existing LEVEL, press: 
 
Levels are displayed according to the following Format: 
  LEVEL ID:  Any number between 0 and 99.  This will be the number used to select the level when 
activating / deactivating it, or when editing or deleting it. 
  Gas Conc: The concentration setting, in ppb, for the O
3
 generator to produce. 
  Status Block Setting: This will be displayed as two pairs of 1-digit numbers.   
  The First pair corresponds to Status Block 1.   
  The Second pair corresponds to Status Block 2. 
 
n each case:  
  The left digit will be a number between 1 and 4 representing the binary setting of bits 1 through 4 
and;  
  The right digit will be a number between 1 and 4 representing the binary setting of bits 5 through 8.  
 
Figure 6-5: 
LEADS Level Display Format 
07223B DCN6378