Teledyne MODEL T703 Manual Do Utilizador

Página de 287
Teledyne API T703 Calibrator Operation Manual 
Operating the T703 Calibrator 
 
103 
6.12.6. CONFIGURING LEVEL STATUS BLOCKS 
There are two STATUS BLOCKS associated with LEADS LEVELS.   
  BLOCK 1: This block corresponds to the physical CONTROL OUTPUT connections located on the back 
  BLOCK 2: The second status block does not correspond to any physical output but is used to 
communicate status over the serial data port 
To configure the either of the STATUS BLOCKS, press: 
SETUP X.X 
[LEVEL ID] 
)[Gas/Conc.],[Status Block Set’g]
PREV NEXT
INS
DEL EDIT
PRNT
EXIT
SETUP X.X    LEVEL NUMBER:12
<SET SET>
EDIT
EXIT
SETUP X.X    STATUS BLOCK 2:DISABLED
<SET SET>
EDIT
EXIT
Continue pressing SET> until Desired 
Status Block is reached
SETUP X.X     STATUS BLOCK 2:[0]0000000
<CH
CH>
  [0]
ENTER EXIT
EXIT discards the 
new setting
ENTR accepts the 
new setting
SETUP X.X     STATUS BLOCK 2:OFF
OFF
ENTER EXIT
 Turns the CC 
input ON/OFF
EXIT discards the 
new setting
ENTR accepts the 
new setting
 Toggle to turn the selected bit ON/OFF (0 or 1).  
Each bit shown on the display represents one of the control 
output pins located on the back of the calibrator (see Figure 3-2),
The left most bit is Bit 1, the next bit to the right, bit 2, 
progressing rightward to bit 8 
(see Figure 3-8 for connector pin assignments)
Moves the 
cursor one 
character left or 
right.
STANDBY            ACT =STANDBY
<TST
TST>
GEN STBY SEQ
SETUP
Make sure that the calibrator 
is in standby mode.
SETUP X.X        
PRIMARY SETUP MENU
O3
LEVL
SEQ
CFG
CLK PASS
MORE
EXIT
Toggle to scroll to the  
LEVEL for editing.
See 
Figure 6-5
 
07223B DCN6378