Operating The T703 over the Serial I/O Ports 
 
111 
7.1.6. MACHINE ID  
Each type of Teledyne API calibrator is configured with a default ID code.  The default ID code for all T703 
calibrators is 703.  The ID number is only important if more than one calibrator is connected to the same 
communications channel such as when several calibrators are on the same Ethernet LAN, in a RS-232 multidrop 
chain (See Section 7.3) or operating over a RS-485 network (See Section 7.4).  If two calibrators of the same 
model type are used on one channel, their ID codes must differ; opt to change one or both. 
To edit the instrument’s ID code, press: 
STANDBY            ACT =STANDBY
<TST
TST>
GEN STBY SEQ
SETUP
SETUP X.X        
PRIMARY SETUP MENU
O3
SEQ CFG
CLK PASS
MORE
EXIT
SETUP X.X        
SECONDARY SETUP MENU
COMM
VARS DIAG
EXIT
SETUP X.X        
COMMUNICATIONS MENU
ID    INET    COM1 COM2
EXIT
SETUP X.X        
MACHINE ID:703 ID
0
7
0
0
ENTR
EXIT
Toggle to cycle 
through the available 
character set:  0-9
EXIT discards the new 
setting
ENTR accepts the 
new setting
 
 
The ID number is only important if more than one calibrator is connected to the same communications channel 
(e.g., a multi-drop setup).  Different models of Teledyne API’s calibrators have different default ID numbers, but if 
two calibrators of the same model type are used on one channel (for example, two T703’s), the ID of one 
instrument needs to be changed.   
The ID can also be used for to identify any one of several calibrators attached to the same network but situated 
in different physical locations. 
 