T703 Calibration and Verification
135
8.1.5. O
3
PHOTOMETER BACKPRESSURE CALIBRATION
Any time there is a pneumatic configuration change, there is risk of impacting the internal measure/reference
pressure. To compensate for this, a backpressure compensation calibration is required each time. Set the
calibrator to generate ozone at the flow rate intended for operation. While the instrument is generating ozone, go
to the SETUP>MORE>DIAG>929>…>BACKPRESSURE COMPENSATION menu and press ENTR, shown in
the following illustration, to initiate the calibration; the operation will take a few minutes:
pressure. To compensate for this, a backpressure compensation calibration is required each time. Set the
calibrator to generate ozone at the flow rate intended for operation. While the instrument is generating ozone, go
to the SETUP>MORE>DIAG>929>…>BACKPRESSURE COMPENSATION menu and press ENTR, shown in
the following illustration, to initiate the calibration; the operation will take a few minutes:
GENERATE A-CAL=0.000 LPM
<TST
TST>
GEN STBY SEQ
SETUP
Make sure that the T700 is
generating ozone at the
intended operational flow rate.
SETUP X.X
PRIMARY SETUP MENU
GAS
SEQ
CFG
CLK PASS
MORE
EXIT
SETUP X.X
SECONDARY SETUP MENU
COMM FLOW
VARS
DIAG
EXIT
SETUP X.X
ENTER PASSWORD
8
1
8
ENTR
EXIT
Toggle these buttons to
enter the correct
PASSWORD - 929
DIAG
SIGNAL I/O
PREV
NEXT
ENTR
EXIT
Continue pressing NEXT until ...
DIAG
BACKPRESSURE COMPENSATION
PREV
NEXT
ENTR
EXIT
07223B DCN6378