Teledyne MODEL T703 Manual Do Utilizador

Página de 287
Teledyne API T703 Calibrator Operation Manual 
T703 Calibration and Verification 
 
137 
8.2.2. VIEWING O
3
 GENERATOR CALIBRATION POINTS 
To view these calibration points, press: 
SETUP X.X        SECONDARY SETUP MENU
COMM VARS DIAG
EXIT
Make sure that the 
calibrator is in standby 
mode.
SETUP X.X      ENTER PASSWORD
8
1
8
ENTR
EXIT
Toggle to enter the correct
PASSWORD
DIAG                         SIGNAL I/O
PREV NEXT
ENTR
EXIT
DIAG                   O3 GEN CALIBRATION
PREV NEXT
ENTR
EXIT
Continue pressing NEXT  until ...
DIAG                   O3 GEN CALIBRATION
CAL
PNTS
EXIT
Toggle to move between 
calibration points
DIAG O3GEN         1)  500 MV, 5.0 MIN, ON
PREV NEXT
INS
DEL EDIT PRNT
EXIT
STANDBY             ACT =STANDBY
<TST
TST> GEN STBY SEQ
SETUP
SETUP X.X        PRIMARY SETUP MENU
O3
SEQ CFG
CLK PASS MORE
EXIT
 
 
07223B DCN6378