T703 Calibration and Verification
137
8.2.2. VIEWING O
3
GENERATOR CALIBRATION POINTS
To view these calibration points, press:
SETUP X.X SECONDARY SETUP MENU
COMM VARS DIAG
EXIT
Make sure that the
calibrator is in standby
mode.
SETUP X.X ENTER PASSWORD
8
1
8
ENTR
EXIT
Toggle to enter the correct
PASSWORD
DIAG SIGNAL I/O
PREV NEXT
ENTR
EXIT
DIAG O3 GEN CALIBRATION
PREV NEXT
ENTR
EXIT
Continue pressing NEXT until ...
DIAG O3 GEN CALIBRATION
CAL
PNTS
EXIT
Toggle to move between
calibration points
DIAG O3GEN 1) 500 MV, 5.0 MIN, ON
PREV NEXT
INS
DEL EDIT PRNT
EXIT
STANDBY ACT =STANDBY
<TST
TST> GEN STBY SEQ
SETUP
SETUP X.X PRIMARY SETUP MENU
O3
SEQ CFG
CLK PASS MORE
EXIT