Teledyne API T703 Calibrator Operation Manual 
Theory of Operation 
 
161 
RELAY PCA
SENSOR SUITES
LOGIC DEVICES
(e.g. CPU, I
2
C bus, 
Motherboard, etc.)
ON / OFF 
SWITCH
PS 2 
(+12 VDC)
Photometer 
M/R valve
Cooling 
Fan
PS 1 
ANALOG SENSORS
O
3
Generator 
Reference detector,
Photometer UV 
Detector
Pre-Amplifiers 
& Amplifiers
Sensor Control 
& I/O Logic
Solenoid 
Drivers
 KEY
  AC POWER
  DC POWER
AC 
POWER IN
+5 VDC
±15 VDC
O
3
Generator UV 
Lamp Xfromer
Photometer 
UV Lamp P/S
GAS 
PRESSURE 
SENSORS
GAS 
TEMPERATURE 
SENSORS
O
3
Generator UV 
Lamp P/S
O
3
Generator
UV Lamp
Photometer 
Pump 
Controlled 
via I
2
C
DRY AIR  
Pump
AC 
Relay 
 
Figure 9-7: 
T703 Power Distribution Block diagram 
9.2.6. AC POWER CONFIGURATION 
The E-Series digital electronic systems will operate with any of the specified power regimes.  As long as 
instrument is connected to 100-120 VAC or 220-240 VAC at either 50 or 60 Hz it will turn on and after about 30 
seconds show a front panel display.  Internally, the status LEDs located on the Relay PCA, Motherboard and 
CPU should turn on as soon as the power is supplied.   
On the other hand, some of the calibrator’s the dry air pump must be properly configured for the type of power 
being supplied to the instrument.  Figure 2-3 shows the location of the Pump AC Configuration jumper. 
07223B DCN6378