Teledyne MODEL T703 Manual Do Utilizador

Página de 287
Theory of Operation 
Teledyne API T703 Calibrator Operation Manual 
162  
 
     
JP7
 
Pump 
Configuration 
 
Figure 9-8: 
Location of the AC Configuration Jumper for the Dry Air Pump 
 
9.2.6.1. AC configuration – Internal Pump (JP7) 
AC power configuration for the internal dry air pump is set using Jumper set JP7.   
Table 9-2:  AC Power Configuration for Internal Pumps (JP7) 
LINE 
POWER 
LINE 
FREQUENCY 
JUMPER 
COLOR 
FUNCTION 
JUMPER 
BETWEEN 
PINS 
Connects pump pin 3 to 110 / 115 VAC power line 
2 to 7 
Connects pump pin 3 to 110 / 115 VAC power line 
3 to 8 
60 HZ 
WHITE 
Connects pump pins 2 & 4  to Neutral 
4 to 9 
Connects pump pin 3 to 110 / 115 VAC power line 
2 to 7 
Connects pump pin 3 to 110 / 115 VAC power line 
3 to 8 
110VAC 
 115 VAC 
50 HZ
BLACK 
Connects pump pins 2 & 4  to Neutral 
4 to 9 
Connects pump pins 3  and 4 together 
1 to 6 
60 HZ 
BROWN 
Connects pump pin 1 to 220 / 240VAC power line 
3 to 8 
Connects pump pins 3  and 4 together 
1 to 6 
220VAC 
240 VAC 
50 HZ
BLUE 
Connects pump pin 1 to 220 / 240VAC power line 
3 to 8 
1
 A jumper between pins 5 and 10 may be present on the jumper plug assembly, but is only functional on the T300 and 
has no function on the Models T700 or T703. 
 
07223B DCN6378