Teledyne API T703 Calibrator Operation Manual 
General Troubleshooting & Repair 
  
201 
A voltmeter should be used to verify that the DC voltages are correct per the values in the table below, and an 
oscilloscope, in AC mode, with band limiting turned on, can be used to evaluate if the supplies are producing 
excessive noise (> 100 mV p-p).   
Table 11-7:  DC Power Supply Acceptable Levels 
CHECK RELAY PCA TEST POINTS 
FROM TEST POINT 
TO TEST POINT 
POWER 
SUPPLY 
ASSY 
VOLTAGE 
NAME # NAME # 
MIN V 
MAX V 
PS1 +5  Dgnd 1  +5  2 4.8 5.25 
PS1 +15  Agnd 3  +15  4 13.5 16V 
PS1 -15  Agnd 3 -15V 5 -14V -16V 
PS1 Agnd  Agnd  3  Dgnd  1 -0.05 0.05 
PS1 Chassis  Dgnd  1  Chassis N/A -0.05  0.05 
PS2 +12 +12V 
Ret 
6 +12V 7 11.75 12.5 
PS2 
+12 V ret 
+12V Ret 
Dgnd 
-0.05 
0.05 
 
 
11.4.4. I
2
C BUS 
Operation of the I
2
C bus can be verified by observing the behavior of D1 on the relay PCA & D2 on the Valve 
Driver PCA. Assuming that the DC power supplies are operating properly, the I
2
C bus is operating properly if D1 
on the relay PCA and D2 of the Valve Driver PCA are flashing 
There is a problem with the I
2
C bus if both D1 on the relay PCA and D2 of the Valve Driver PCA are ON/OFF 
constantly. 
 
11.4.5. TOUCHSCREEN INTERFACE 
Verify the functioning of the touch screen by observing the display when pressing a touch-screen control button. 
Assuming that there are no wiring problems and that the DC power supplies are operating properly, but pressing 
a control button on the touch screen does not change the display, any of the following may be the problem: 
  The touch-screen controller may be malfunctioning.  
   The internal USB bus may be malfunctioning.  
You can verify this failure by logging on to the instrument using APICOM or a terminal program. If the analyzer 
responds to remote commands and the display changes accordingly, the touch-screen interface may be faulty.  
11.4.6. LCD DISPLAY MODULE 
Verify the functioning of the front panel display by observing it when power is applied to the instrument. 
Assuming that there are no wiring problems and that the DC power supplies are operating properly, the display 
screen should light and show the splash screen and other indications of its state as the CPU goes through its 
initialization process. 
 