Teledyne MODEL T703 Manual Do Utilizador

Página de 287
General Troubleshooting & Repair 
Teledyne API T703 Calibrator Operation Manual 
202  
 
11.4.7. RELAY PCA 
The Relay PCA can be most easily checked by observing the condition of the status LEDs located along its 
upper edge (see Section 11.3.2 and Figure 11-3: Relay PCA Status LEDS Used for Troubleshooting), and using 
the SIGNAL I/O submenu under the DIAG menu (see Section 11.1.3) to toggle each LED ON or OFF.  
If D1 on the Relay PCA is flashing and the status indicator for the output in question (Pump power, Heater 
power, Valve Drive, etc.) toggles properly using the Signal I/O function, then the associated control device on the 
Relay PCA is bad.  Several of the control devices are in sockets and can be easily replaced.  The table below 
lists the control device associated with a particular function. 
Table 11-8:  Relay PCA Control Devices 
FUNCTION 
CONTROL 
DEVICE 
IN SOCKET 
UV Lamp Heater 
Q2 
No 
O
3
 Gen Heater 
Q3 
No 
All Valves 
U5 
Yes 
AC Dry air Pump 
K1 
No 
DC Photometer Pump 
U1 
No 
 
 
 
11.4.8. PHOTOMETER O
3
 GENERATOR PRESSURE /FLOW SENSOR 
ASSEMBLY 
This assembly is only present in calibrators with O
3
 generator and/or photometer options installed.  The 
pressure/flow sensor PCA, located at the rear of the instrument between the O
3
 generator and the photometer 
pump (see Figure 3-5) can be checked with a Voltmeter.  The following procedure assumes that the wiring is 
intact and that the motherboard as well as the power supplies are operating properly: 
BASIC PCA OPERATION: 
  Measure the voltage across C1 it should be 5 VDC ± 0.25 VDC.  If not then the board is bad 
  Measure the voltage between TP2 and TP1 C1 it should be 1o VDC ± 0.25 VDC.  If not then the board is 
bad. 
07223B DCN6378