3.2. ELECTRICAL CONNECTIONS 
NOTE  
To maintain compliance with EMC standards, it is required that the cable length be no greater than 3 meters for 
all I/O connections, which include Analog In, Analog Out, Status Out, Control In, Ethernet/LAN, USB, RS-232, 
and RS-485.   
3.2.1. POWER CONNECTION 
Verify the correct line voltage and frequency configuration on the serial number tag on the rear panel of the 
T703. 
Attach the power cord to the calibrator and plug it into a power outlet capable of carrying at least 10 A current at 
your AC voltage and that it is equipped with a functioning earth ground. 
 
WARNING 
HIGH VOLTAGES ARE PRESENT INSIDE THE CALIBRATORS CASE 
POWER CONNECTION MUST HAVE FUNCTIONING GROUND CONNECTION. 
DO NOT DEFEAT THE GROUND WIRE ON POWER PLUG. 
TURN OFF CALIBRATOR POWER BEFORE DISCONNECTING OR  
CONNECTING ELECTRICAL SUBASSEMBLIES.
 
 
 
CAUTION – AVOID PERSONAL INJURY 
DO NOT LOOK AT THE PHOTOMETER UV LAMP. 
UV LIGHT CAN CAUSE EYE DAMAGE. 
ALWAYS WEAR GLASSES MADE FROM SAFETY UV FILTERING GLASS  
(PLASTIC GLASSES ARE INADEQUATE). 
 
3.2.2. ANALOG OUTPUT TEST CHANNEL CONNECTIONS 
The T703 is equipped with an analog output channel accessible through a connector on the back panel of the 
instrument.  The standard configuration for this output is 0-5 VDC.  It can be set by the user to output one of a 
variety of diagnostic test functions (see Section 6.9.) 
To access these signals attach a strip chart recorder and/or data-logger to the appropriate analog output 
connections on the rear panel of the calibrator.   
Pin-outs for the analog output connector at the rear panel of the instrument are: 
 
ANALOG OUT 
+    –
 
Figure 3-7: 
T703 the TEST CHANNEL Connector 
07223B DCN6378