Getting Started 
 
35 
3.2.5. CONNECTING THE CONTROL OUTPUTS 
The calibrator is equipped with 12 opto-isolated, digital control outputs.  These outputs are activated by the 
T703’s user-programmable, calibration sequences (see Section 6.5.1.6 for instructions on assigning the control 
outputs to specific calibration sequences) 
These outputs may be used to interface with devices that accept logic-level digital inputs, such as programmable 
logic controllers (PLCs), dataloggers, or digital relays/valve drivers.   
They are accessed via a 14-pin connector on the calibrator’s rear panel (see Figure 3-4). 
 
CONTROL OUTPUTS
     1      2      3      4       5     6     
 
7      
 
8      9     
 
10    
 
11    12     
 
E      
 
Figure 3-10: 
T703 Digital Control Output Connector 
 
NOTE 
Most PLCs have internal provisions for limiting the current the input will draw.  When connecting to a 
unit that does not have this feature, external resistors must be used to limit the current through the 
individual transistor outputs to ≤50mA (120 Ω for 5V supply).   
 
The pin assignments for the control outputs are: 
Table 3-5:  T703 Control Input Pin Assignments 
PIN # 
STATUS DEFINITION 
CONDITION 
1 - 12 
Outputs 1 through 12 respectively   Closed if the sequence or sequence step activating output is operating 
Emitter BUSS 
The emitters of the transistors on pins 1 to 8 are bussed together. 
 
Digital Ground 
The ground level from the calibrator’s internal DC power supplies. 
 