Operating the T703 Calibrator 
Teledyne API T703 Calibrator Operation Manual 
82    
6.9.1.2. Selecting a TEST Channel Function to Output 
The Test Functions available to be reported are: 
Table 6-8:  Test Channels Functions Available on the T703’s Analog Output 
TEST CHANNEL 
DESCRIPTION 
ZERO 
FULL SCALE 
NONE 
TEST CHANNEL IS TURNED OFF 
O3 PHOTO MEAS 
The raw output of the photometer during its 
measure cycle 
0 mV 
5000 mV* 
O3 PHOTO REF 
The raw output of the photometer during its 
reference cycle 
0 mV 
5000 mV* 
O3 GEN REF 
The raw output of the O
3
 generator’s 
reference detector 
0 mV 
5000 mV* 
OUTPUT FLOW 
The gas flow being output through the CAL 
GAS outlets on the back of the instrument 
0 cm
3
/min 5,000 
cm
3
/min 
REGULATOR PRESSURE 
The gas pressure measured by the O
3
 
generator pressure sensor 
0 PSIG 
105 PSIG 
SAMPLE PRESSURE 
The pressure of gas in the photometer 
absorption tube 
0 "Hg 
40 "Hg-In-A 
SAMPLE FLOW 
The gas flow rate through the photometer 
0 cm
3
/min 1000 
cc
3
/min 
SAMPLE TEMP 
The temperature of gas in the photometer 
absorption tube 
C 70 
C 
PHOTO LAMP TEMP 
The temperature of the photometer UV lamp 
0 C
C 70 
C 
O3 LAMP TEMP 
The temperature of the O
3
 generator’s UV 
lamp 
0 mV 
5000 mV 
CHASSIS TEMP 
The temperature inside the T703’s chassis 
(same as BOX TEMP
C 70 
C 
O3 PHOTO CONC 
The current concentration of O
3
 being 
measured by the photometer. 
C 
1 ppm 
 
Once a function is selected, the instrument not only begins to output a signal on the analog output, but also adds 
TEST to the list of Test Functions viewable via the Front Panel Display.  