Operating the T703 Calibrator 
Teledyne API T703 Calibrator Operation Manual 
84    
6.9.1.3. TEST Channel Voltage Range Configuration 
In its standard configuration the analog outputs is set to output a 0 – 5 VDC signals.  Several other output 
ranges are available (see Table 7-5).  Each range has is usable from -5% to + 5% of the rated span.  
 
Table 6-9:  Analog Output Voltage Range Min/Max 
RANGE SPAN 
MINIMUM OUTPUT 
MAXIMUM OUTPUT 
0-100 mVDC 
-5 mVDC 
105 mVDC 
0-1 VDC 
-0.05 VDC 
1.05 VDC 
0-5 VDC 
-0.25 VDC 
5.25 VDC 
0-10 VDC 
-0.5 VDC 
10.5 VDC 
The default offset for all ranges is 0 VDC. 
 
To change the output range, press, 
 
DIAG AIO           AOUTS CALIBRATED: NO
  SET>
 CAL
EXIT
From the
AIO CONFIGURATION SUBMENU
(See Section 6.9.1.1)
DIAG              
ANALOG I/O CONFIGURATION             
PREV   NEXT  
 ENTR
EXIT
Continue pressing SET> until you reach the 
output to be configured
DIAG AIO        TEST_OUTPUT: 5V, OVR, NOCAL
<SET SET>
  EDIT
EXIT
DIAG AIO        TEST_OUTPUT: RANGE: 5V
0.1V    1V
  5V   10V
ENTR EXIT
These buttons 
set the signal 
level and type 
of the selected 
channel
Pressing ENTR records 
the new setting and 
returns to the previous 
menu.
Pressing EXIT ignores the 
new setting and returns to 
the previous menu.
 
 