Teledyne MODEL T703 Manual Do Utilizador
Operating the T703 Calibrator
Teledyne API T703 Calibrator Operation Manual
84
6.9.1.3. TEST Channel Voltage Range Configuration
In its standard configuration the analog outputs is set to output a 0 – 5 VDC signals. Several other output
ranges are available (see Table 7-5). Each range has is usable from -5% to + 5% of the rated span.
ranges are available (see Table 7-5). Each range has is usable from -5% to + 5% of the rated span.
Table 6-9: Analog Output Voltage Range Min/Max
RANGE SPAN
MINIMUM OUTPUT
MAXIMUM OUTPUT
0-100 mVDC
-5 mVDC
105 mVDC
0-1 VDC
-0.05 VDC
1.05 VDC
0-5 VDC
-0.25 VDC
5.25 VDC
0-10 VDC
-0.5 VDC
10.5 VDC
The default offset for all ranges is 0 VDC.
To change the output range, press,
DIAG AIO AOUTS CALIBRATED: NO
SET>
CAL
EXIT
From the
AIO CONFIGURATION SUBMENU
(See Section 6.9.1.1)
DIAG
ANALOG I/O CONFIGURATION
PREV NEXT
ENTR
EXIT
Continue pressing SET> until you reach the
output to be configured
DIAG AIO TEST_OUTPUT: 5V, OVR, NOCAL
<SET SET>
EDIT
EXIT
DIAG AIO TEST_OUTPUT: RANGE: 5V
0.1V 1V
5V 10V
ENTR EXIT
These buttons
set the signal
level and type
of the selected
channel
Pressing ENTR records
the new setting and
returns to the previous
menu.
Pressing EXIT ignores the
new setting and returns to
the previous menu.
07223B DCN6378