Operating the T703 Calibrator 
Teledyne API T703 Calibrator Operation Manual 
88    
6.9.2.1. Enabling or Disabling the TEST CHANNEL Auto-Cal Feature 
To enable or disable the Auto-Cal feature for the TEST CHANNEL, press. 
From the
AIO CONFIGURATION SUBMENU
(See Section 6.9.1.1.)
DIAG              
ANALOG I/O CONFIGURATION             
PREV   NEXT  
 ENTR
EXIT
DIAG AIO        TEST_OUTPUT: 5V, OVR, NOCAL
<SET SET>
EDIT
EXIT
DIAG AIO        TEST_OUTPUT: AUTO CAL.:ON
<SET SET>
EDIT
EXIT
DIAG AIO        TEST_OUTPUT: AUTO CAL.:ON
 ON
ENTR EXIT
Toggle to turn  
AUTO CAL 
ON or OFF 
(OFF = manual 
calibration mode).
ENTR accepts 
the new setting.
EXIT ignores the 
new setting
Continue pressing SET> until ...
DIAG AIO        TEST_OUTPUT: RANGE: 5V
SET>
  EDIT
EXIT
DIAG AIO        TEST_OUTPUT: AUTO CAL.:OFF
 OFF
ENTR EXIT
Continue pressing SET> until you reach the 
output to be configured
DIAG AIO           AOUTS CALIBRATED: NO
  SET>
 CAL
EXIT
NOTE: 
TEST CHANNELS
configured for 0.1V full 
scale should always be 
calibrated manually.
 
6.9.2.2. Automatic TEST Channel Calibration 
To calibrate the outputs as a group with the AOUTS CALIBRATION command, press: 
NOTE 
Before performing this procedure, make sure that the AUTO CAL feature is turned OFF for CONC_OUT_1 
and CONC_OUT_2, 
Make sure that the AUTO CAL feature is turned ON for the TEST CHANNEL (See Section 6.9.2.1) 
 