Teledyne API – Model T300/T300M CO Analyzer 
Setup Menu 
137 
5.9.4. ELECTRICAL TEST 
The electrical test function creates a current, which substitutes the PMT signal, and 
feeds it into the preamplifier board. This signal is generated by circuitry on the pre-
amplifier board itself and tests the filtering and amplification functions of that assembly 
along with the A/D converter on the motherboard. It does not test the PMT itself. (See 
also Section 
 
 
 
5.9.5. DARK CALIBRATION 
The dark calibration test interrupts the signal path between the IR photo-detector and the 
remainder of the sync/demod board circuitry.  This allows the instrument to compensate 
for any voltage levels inherent in the sync/demod circuitry that might affect the 
calculation of CO concentration. For details see Section 9.6.1. 
5.9.6. PRESSURE CALIBRATION 
A sensor at the exit of the sample chamber continuously measures the pressure of the 
sample gas. The data are used to compensate the final CO concentration calculation for 
changes in atmospheric pressure and are stored in the CPU’s memory as the test function 
PRES (also viewable via the front panel). For details see Section 9.6.2. 