Teledyne t320u Manual Do Utilizador
Setup Menu
Teledyne API – Model T300/T300M CO Analyzer
136
5.9.3.11. ANALOG INPUTS (XIN1…XIN8) OPTION CONFIGURATION
To configure the analyzer’s external analog inputs option, define for each channel:
gain (number of units represented by 1 volt)
offset
offset
(volts)
engineering units to be represented in volts (each press of the touchscreen
button scrolls the list of alphanumeric characters from A-Z and 0-9)
whether to display the channel in the Test functions
To access and adjust settings for the external Analog Inputs option channels press:
DIAG AIO
XIN1 GAIN:1.00V/V
SET> EDIT
EXIT
DIAG
ANALOG I / O CONFIGURATION
PREV
NEXT
ENTR
EXIT
DIAG AIO
AOUTS CALIBRATED: NO
< SET SET> CAL
EXIT
DIAG AIO
XIN1 GAIN:1.00V/V
+ 0 0 1 .0 0 ENTR EXIT
Press to change
Gain value
DIAG AIO
XIN1:1.00,0.00,V,OFF
< SET SET> EDIT
EXIT
Press SET> to scroll to the first
channel. Continue pressing SET>
to view each of 8 channels.
channel. Continue pressing SET>
to view each of 8 channels.
Pressing ENTR records the new setting
and returns to the previous menu.
Pressing EXIT ignores the new setting and
returns to the previous menu.
Press EDIT at any channel
to to change Gain, Offset,
Units and whether to display
the channel in the Test
functions (OFF/ON).
to to change Gain, Offset,
Units and whether to display
the channel in the Test
functions (OFF/ON).
DIAG AIO
XIN1 OFFSET:0.00V
< SET
SET> EDIT
EXIT
DIAG AIO
XIN1 UNITS:V
< SET SET> EDIT
EXIT
DIAG AIO
XIN1 DISPLAY:OFF
< SET EDIT
EXIT
Figure 5-7. DIAG – Analog Inputs (Option) Configuration Menu
06864B DCN6314