Teledyne t320u Manual Do Utilizador

Página de 426
Data Acquisition System (DAS) and APICOM 
Teledyne API – Model T300/T300M CO Analyzer 
188 
7.2.2. DAS CONFIGURATION USING TERMINAL EMULATION PROGRAMS 
Although Teledyne API recommends the use of APICOM, the DAS can also be 
accessed and configured through a terminal emulation program such as HyperTerminal 
(see example in Figure 7-4)
To do this: 
  All configuration commands must be created and edited off line (e.g. cut & pasted in 
from a text file or word processor) following a strict syntax (see below for example). 
  The script is then uploaded via the instrument’s RS-232 port(s). 
 
Figure 7-4: 
DAS Configuration Through a Terminal Emulation Program 
Both of the above steps are best started by: 
1. Downloading 
the 
default DAS configuration. 
2.  Getting familiar with its command structure and syntax conventions. 
3.  Altering a copy of the original file offline. 
4.  Uploading the new configuration into the analyzer. 
I
MPORTANT
 
I
MPACT ON 
R
EADINGS OR 
D
ATA
 
Whereas the editing, adding and deleting of DAS channels and 
parameters of one channel through the front-panel control buttons can 
be done without affecting the other channels, uploading a DAS 
configuration script to the analyzer through its communication ports will 
erase all data, parameters and channels by replacing them with the new 
DAS configuration. Backup of data and the original DAS configuration is 
advised before attempting any DAS changes. 
Refer to Section 8.2.1 for details on remote access to and from the T300/T300M 
Analyzer via the instrument’s COMM ports.  
06864B DCN6314