Teledyne t320u Manual Do Utilizador

Página de 426
Remote Operation 
Teledyne API – Model T300/T300M CO Analyzer 
190 
8.2. INTERACTIVE MODE 
Interactive mode is used with a terminal emulation programs or a “dumb” computer 
terminal.  
8.2.1. REMOTE CONTROL VIA A TERMINAL EMULATION PROGRAM 
Start a terminal emulation program such as HyperTerminal. All configuration 
commands must be created following a strict syntax or be pasted in from an existing text 
file, which was edited offline and then uploaded through a specific transfer procedure. 
The commands that are used to operate the analyzer in this mode are listed in Table 8-1 
and in Appendix A. 
8.2.1.1. HELP COMMANDS IN INTERACTIVE MODE 
Table 8-1:  Interactive Mode Software Commands 
COMMAND Function 
Control-T 
Switches the analyzer to terminal mode (echo, edit).  If mode flags 1 & 2 are OFF, the 
interface can be used in interactive mode with a terminal emulation program. 
Control-C 
Switches the analyzer to computer mode (no echo, no edit). 
CR(carriage return) 
A carriage return is required after each command line is typed into the terminal/computer.  
The command will not be sent to the analyzer to be executed until this is done.  On 
personal computers, this is achieved by pressing the ENTER button. 
BS (backspace) 
Erases one character to the left of the cursor location. 
ESC(escape) 
Erases the entire command line. 
?[ID] CR 
This command prints a complete list of available commands along with the definitions of 
their functionality to the display device of the terminal or computer being used.  The ID 
number of the analyzer is only necessary if multiple analyzers are on the same 
communications line, such as the multi-drop setup. 
8.2.1.2. COMMAND SYNTAX 
Commands are not case-sensitive and all arguments within one command (i.e.  ID 
numbers, key words, data values, etc.) must be separated with a space character. 
All Commands follow the syntax: 
X [ID] COMMAND <CR> 
Where  
X  
is the command type (one letter) that defines the type of command.  
Allowed designators are listed in Appendix A-6. 
[ID] 
is the machine identification number (Section 5.7.1).  Example: the 
Command “? 700” followed by a carriage return would print the list of 
available commands for the revision of software currently installed in the 
instrument assigned ID Number 700. 
COMMAND  is the command designator: This string is the name of the command being 
issued (LIST, ABORT, NAME, EXIT, etc.).  Some commands may have 
additional arguments that define how the command is to be executed.  
Press ? <CR> or refer to Appendix A-6 for a list of available command 
designators. 
<CR>  
is a carriage return.  All commands must be terminated by a carriage 
return (usually achieved by pressing the ENTER button on a computer). 
06864B DCN6314