Calibration Procedures 
Teledyne API – Model T300/T300M CO Analyzer 
222 
9.6.3. FLOW CALIBRATION 
The flow calibration allows the user to adjust the values of the sample flow rates as they 
are displayed on the front panel and reported through COMM ports to match the actual 
flow rate measured at the sample inlet.  This does not change the hardware measurement 
of the flow sensors, only the software-calculated values.   
To carry out this adjustment, connect an external, sufficiently accurate flow meter to the 
sample inlet (see Section 11.3.4 for more details).  Once the flow meter is attached and 
is measuring actual gas flow, press: 
 
 
See also Section 
 