Teledyne t320u Manual Do Utilizador

Página de 426
Teledyne API – Model T300/T300M CO Analyzer 
Calibration Procedures 
223 
9.7.  CALIBRATION OF OPTIONAL SENSORS 
This section provides the calibration setup and procedures for the O
2
 Sensor and the CO
2
 
Sensor options. 
9.7.1. O
2
 SENSOR CALIBRATION 
Presented here are first the setup and then the calibration steps for the O
2
 Sensor. 
9.7.1.1. O2 PNEUMATICS CONNECTIONS 
The pneumatic connections for calibrating are as follows:  
 
Source of
SAMPLE GAS
Removed during 
calibration
 SAMPLE
 EXHAUST
PUMP
3-way
Valve
Manual 
Control Valve
VENT 
here if input 
is pressurized
Instrument 
Chassis
 
Figure 9-7: 
O
2
 Sensor Calibration Set Up 
O
2
 SENSOR ZERO GAS: Teledyne API recommends using pure N
2
 when calibration 
the zero point of your O
2
 sensor option. 
O
2
 SENSOR SPAN GAS: Teledyne API recommends using 20.8% O
2
 in N
2
 when 
calibration the span point of your O
2
06864B DCN6314