Teledyne t320u Manual Do Utilizador
Teledyne API – Model T300/T300M CO Analyzer
Specifications and Approvals
33
Table 2-3:
CO
2
Sensor Option Specifications
2.2. EPA EQUIVALENCY DESIGNATION
Note
T300M: EPA equivalency does not apply to this model.
Teledyne API’s Model T300, Carbon Monoxide Analyzer, is designated as Reference
Method Number RFCA-1093-093 as defined in 40 CFR Part 53, when operated under
the following conditions:
Method Number RFCA-1093-093 as defined in 40 CFR Part 53, when operated under
the following conditions:
Range: Any range from 10 ppm to 50 ppm
Ambient temperature range of 10 to 40C
Sample filter: Equipped with 5-micron PTFE filter element in the internal filter
Ambient temperature range of 10 to 40C
Sample filter: Equipped with 5-micron PTFE filter element in the internal filter
assembly
Software
settings:
Dilution factor
1.0
AutoCal
ON or OFF
Dynamic Zero
ON or OFF
Dynamic Span
OFF
Dual Range
ON or OFF
Auto Range
ON or OFF
Temp/Pres Compensation
ON
Parameter Description
Ranges
0-1% to 0-20% user selectable. Dual ranges and auto-ranging supported.
Zero Noise
1
<0.02% CO
2
Lower Detectable Limit
2
<0.04%
CO
2
Zero Drift (24 hours)
<± 0.02% CO
2
Zero Drift (7 days)
<± 0.05% CO
2
Span Noise
1
<± 0.1% CO
2
Span Drift (7 days)
<± 0.1% CO
2
Accuracy
<± (0.02% CO
2
+ 2% of reading)
Linearity
<± 0.1% CO
2
Temperature Coefficient
<± 0.01% CO
2
/°C
Rise and Fall Time
<60 seconds to 95%
1
As defined by the USEPA
2
Defined as twice the zero noise level by the USEPA
06864B DCN6314