Specifications and Approvals 
33 
Table 2-3: 
CO
2
 Sensor Option Specifications 
2.2. EPA EQUIVALENCY DESIGNATION 
Note 
T300M: EPA equivalency does not apply to this model. 
Teledyne API’s Model T300, Carbon Monoxide Analyzer, is designated as Reference 
Method Number RFCA-1093-093 as defined in 40 CFR Part 53, when operated under 
the following conditions: 
  Range: Any range from 10 ppm to 50 ppm 
  Ambient temperature range of 10 to 40C  
  Sample filter: Equipped with 5-micron PTFE filter element in the internal filter 
assembly  
 Software 
settings: 
Dilution factor 
1.0 
AutoCal 
ON or OFF 
Dynamic Zero 
ON or OFF 
Dynamic Span 
OFF 
Dual Range 
ON or OFF 
Auto Range 
ON or OFF 
Temp/Pres Compensation 
ON 
Parameter Description 
Ranges 
0-1% to 0-20% user selectable. Dual ranges and auto-ranging supported. 
Zero Noise
<0.02% CO
2
 
Lower Detectable Limit
2
 <0.04% 
CO
2
 
Zero Drift (24 hours) 
<± 0.02% CO
2
 
Zero Drift (7 days) 
<± 0.05% CO
2
 
Span Noise
<± 0.1% CO
2
 
Span Drift (7 days) 
<± 0.1% CO
2
 
Accuracy 
<± (0.02% CO
+ 2% of reading) 
Linearity 
<± 0.1% CO
2
 
Temperature Coefficient 
<± 0.01% CO
2
 /°C 
Rise and Fall Time 
<60 seconds to 95% 
1
 As defined by the USEPA 
2
 Defined as twice the zero noise level by the USEPA
  
06864B DCN6314