Emerson IM-106-5500 Manual Do Utilizador

Página de 108
Instruction Manual
IM-106-5500, Original Issue
August 2005
1-3
CCO 5500
The radiation emitted by the heater is then 'chopped' by a motor-driven disc, 
and focused across the duct by a lens. The chopper disc is driven by a small 
DC motor. The phase and frequency of the chopper disc are monitored by a 
radiation detector to provide a reference signal that is utilized by the signal 
processor unit. 
System calibration is achieved by a small calibration cell containing pure 
reference gas (CO) that can be swung into the sight path by means of a 
'stepper' motor and gearbox assembly to enable continuous calibration 
updates to be maintained. The chopper motor and stepper motor represent 
the only moving components in the entire system. 
A printed circuit board mounted at the front of the unit provides control 
circuitry for the heater, the motors, and the reference wave detector. 
Figure 1-2 illustrates the transmitter and receiver units.
Figure 1-2. Transmitter and 
Receiver Schematic
Infrared Receiver Unit 
The precision and reliability which CO concentration levels can be measured 
governs the performance of the complete instrument. For this reason design 
efforts have been concentrated in producing an extremely simple and robust 
receiver unit. It contains no moving parts, is fully sealed and designed to give 
many years of trouble-free and maintenance-free operation. 
The unit comprises a lens to focus radiation received from across the duct, 
followed by a precision interference filter to limit the wave band of energy 
used. This filter tolerance is strictly controlled since it alone determines the 
instrument scale shape and calibration. The radiation then passes to an 
optical beam-splitter where approximately half the radiation is reflected at 
right angles directly onto a radiation detector. The other half of the radiation is 
transmitted, by the beam-splitter, through a gas cell containing pure reference 
gas (CO) and onto a second radiation detector. 
The detectors used are lithium tantalate pyro-electric detectors, renowned for 
their sensitivity, stability and ability to operate at normal ambient 
temperatures, without the need for cooling. They respond only to changing 
levels of radiation and thus to the chopped radiation from the infrared source 
unit and not to background radiation from the flue or flue gas. The detector 
signals are amplified and fed to the signal processor unit. 
Chopper Blade Detector
Chopper Blade & Motor
Calibrating Cell
Stepper Motor & Gearbox
Reference Detector
Heated Infrared Source
Control Detector
Detector D2
Gas Filter
Detector D1
Beam Splitter
Interference Filter