If the duct is operating and a high opacity may be present, reduce the set 
voltages to 2V rms maximum. This is to prevent saturation should the opacity 
level drop.
Signal Processor Gain 
Adjustment 
After the detector level(s) at the receiver have been optimized if necessary, 
the levels within the micro-processor should be adjusted. This adjustment is 
conducted by means of two trim potentiometers within the signal processor, 
Figure 3-4.
Figure 3-4. Gain Adjustment 
Potentiometers
1. *Set the gain to a minimum by turning the D2 - detector trim 
potentiometer within the signal processor fully CLOCKWISE - it is a 
20-turn pot. 
2. *Enter the DIAGNOSTIC MODE - DETECTOR LEVELS option, and 
display the values of D2 and D1. Turn the trim pot 
COUNTERCLOCKWISE until the D2 level is between 12,000 and 
15,000. Allow time between adjustments for the readings to settle. 